Today 2012-5-19
 
  Technical Articles
  DownLoad
 

HOME > Technologies > Comparison between reference comparison method  BET method

Comparison between reference comparison method  BET method
Compare item
reference comparison method
BET method
Desorbed gas calibration
Compare directly with standard sample peak area with known surface area, no need for real gas desorbed calibration.
No standard sample required, use gas peak area of known set volume to calibrate the testing samples gas volume.
  Principle
Adsorptions peak areas of a sample and standard are proportional to their adsorbed gad volumes, respectively, therefore samples surface are can be obtained through simple comparison.
Based on multi-layer adsorption theory, single-layer saturation adsorption gas volume can be derived by BET equation for surface area calculation. BET method is the most popular nitrogen adsorption surface area testing method.
 Testing gas
                                            Fixed ratio of He to N2(4:1)
High purity N2 and He, experiment need to choose various mix ratio.
 Application range
Applicable samples which have similar adsorption performance to standards, domestically widely used in carbon black industry, not much use abroad.
Applicable to all kinds of samples without micro-pores. This is internationally accepted method for various applications.
  Result
Test accuracy varies depending on different samples
Using standard set volume gas to calibrate nitrogen adsorption volume, this method enjoy high accuracy regardless samples varieties.
Application area
On-site product line monitoring, fast surface area testing.
Research institutes, universities, enterprises for precise surface area testing.
 
Beijing: Room 802,  A Block,No.25,Binhe Road South,Guang'an Men,Beijing,China
Phone: (8610)-68949825 68949817 63326034
  Shanghai: Room 2604, C Block,Shen Han Building,No.1066, Jingshajiang Road, Shanghai, China
Phone: (021)-52657506

Email:sell@jwgb.cn  Beijing ICP 05067832
Copyright: Beijing JWGB Science &Technology Co., Ltd.